Steam boiler and hot water boiler water quality rapid testing instrument and reagent device

The synopsis of project
The ERUN-SP series portable steam boiler and hot water boiler water quality rapid testing instrument and reagent device developed and produced by Erun Environmental Protection Group is designed based on the national standard of 《GB/T 1576-2018 Water quality for industrial boilers》. It can be used for rapid testing of raw water, softened water, desalted water, boiler feed water, make-up water, boiler water, back water, etc. in natural circulation steam boilers and steam water dual-purpose boilers that use external or internal water treatment, through flow and direct flow steam boilers, hot water boilers.
The contents of project

1、List of steam boiler and hot water boiler water quality rapid testing instrument and reagent device:

Serial No.Part NumberNameItemsPictures
1ERUN-SP-M9000Portable multi-parameter water quality testing instrumentphosphate ions、total alkalinity、total hardness、copper、iron、turbidity、PH、dissolved oxygen、conductivity、total dissolved solids、chlorine ions

Boiler feed water and boiler water laboratory benchtop multi-parameter water quality analyzer


2ERUN-ST-OD380Laboratory benchtop infrared spectrophotometeroil

Boiler feed water and boiler water laboratory benchtop infrared spectrophotometer


Note: Analysis parameter items can be flexibly combined and selected according to the customer's own needs.


2、Reagents List of steam boiler and hot water boiler water quality rapid testing instrument and reagent device:

Serial No.

Name

Specifications

Quantity

1

Phenolphthalein indicator solution

10 g/L,Using ethanol as solvent, 500mL/bottle

1

2

Methyl orange indicator liquid

1 g/L,500mL/bottle

1

3

Methyl red-methylene blue mixed indicator solution

500mL/bottle

1

4

Potassium iodate-potassium iodide standard solution

(1mL is equivalent to 1g sulfite), 500mL/bottle

1

5

Hydrochloric acid solution

(1+1),500mL/bottle

1

6

Sodium hydroxide

500g/bottle

1

7

Starch indicator liquid

10g/L,500mL/bottle

1

8

Phosphate reagent

0.01-10.0/10.0-50.0mg/L,100 times/set

1

9

Total alkalinity reagent

0-100/100-1000mg/L,100 times/set

1

10

Total hardness reagent

0.1-5.0/5.0-200.0mg/L,100 times/set

1

11

Copper reagent

0.02-1.0/1.0-10.0mg/L,100 times/set

1

12

Iron reagent

0.0-5.0/5.0-10.0mg/L,50 times/set

1

13

Chlorine ion reagent

0.20-100.0mg/L,50 times/set

1


3、Other parts List of steam boiler and hot water boiler water quality rapid testing instrument and reagent device:

Serial No.

Name

Specifications

Quantity

1

Colorimetric tube

Ф25×95mm

10

2

Cuvette

50×25×70mm

2

3

Lens hood

Black

1

4

Test tube rack

White

2

5

Pipette gun

Minimum graduation 0.1mL

1

6

Acid buret

25mL,white & brown

1

7

Microburette

5mL,Seat type, PTFE cock, grade A

1

8

Erlenmeyer flask

250mL,Transparent glass

1

9

Glass measuring cylinder

100mL,Transparent glass, minimum graduation 1mL

1

10

Volumetric flask

1000mL,White high borosilicate

1

11

Temperature electrode

/

1

12

PH composite electrode

/

1

13

Dissolved oxygen electrode

/

1

14

Conductivity electrode

/

1

15
Chargers and charging cables/1
16USB cable/1
17Instruction Manual/1
18

Product Certificate

/Product Warranty Card

/1


4、Parameters of steam boiler and hot water boiler water quality rapid testing instrument and reagent device:


Serial No.

Items

Measuring range

Maximum Deviation

GB & Measuring Method

1

phosphate ions

0.01-50.00mg/L

0.2 or 5%

GB/T 6913-2023

Phosphorus molybdenum blue colorimetric method

2

total alkalinity

0.05-10.00mmol/L

5%(F.S)

Acid-base titration

3

phenolphthalein alkalinity

0-50mmol/L

/

4

relative alkalinity

0.00-1.00

/

/

5

total hardness

0.01-5.00mmol/L

5%(F.S)

GB/T 6909-2018

Acid Chromium Blue K method

6

copper

0.02-10.00mg/L

5%(F.S)

GB/T 13689-2007

Sodium diethyldithiocarbamate spectrophotometry

7

iron

0.00-10.00mg/L

5%(F.S)

GB/T 14427-2017

1,10-phenanthroline spectrophotometry

8

turbidity

0.00-400.00FTU

≤2%(F.S)

GB/T 12151-2005

Formazine turbidity

9

PH(25℃)

0.00-14.00

±0.1

GB/T 6904-2008

Combined electrode method

10

dissolved oxygen

0.00-20.00 mg/L

±0.2

GB/T 12157-2022

Oxygen electrode method

11

conductivity

(25℃)

0.00 -199.9×103µS/cm

±1.0% (F.S)

GB/T 6908-2018

Potassium chloride electrode method

12

total dissolved solids

0.0~199.9×103mg/L

±0.5% (F.S)

GB/T 14415-2007

Solid guide ratio method

13

oil

0.00~900.00mg/L

RSD<0.6%

GB/T 12152-2007

Infrared photometry

14

sulfite

0.00-100.00mg/L

/

Iodometric method

15

chlorine ions

0.20-100.00mg/L

5%(F.S)

GB/T 29340-2012

Ammonium thiocyanide titration method

Note: Analysis parameter items can be flexibly combined and selected according to the customer's own needs.


5、Scopes of steam boiler and hot water boiler water quality rapid testing instrument and reagent device:

1、Water quality of natural circulation steam boilers and steam-water boilers using external water treatment:

boiler feed water:

turbidity/FTU、total hardness/(mmol/L)、pH(25℃)、conductivity(25℃)/μS/cm)、dissolved oxygen/(mg/L)、oil/(mg/L)、iron/(mg/L)

boiler water:

total alkalinity/(mmol/L)、phenolphthalein alkalinity/(mmol/L)、pH(25℃)、conductivity(25℃)/μS/cm)、total dissolved solids/(mg/L)、phosphate ions/(mg/L)、sulfite/(mg/L)、relative alkalinity

2、Water quality of natural circulation steam boilers and steam-water boilers using in-pot water treatment:

boiler feed water:

turbidity/FTU、total hardness/(mmol/L)、pH(25℃)、oil/(mg/L)、iron/(mg/L)

boiler water:

total alkalinity/(mmol/L)、phenolphthalein alkalinity/(mmol/L)、pH(25℃)、conductivity(25℃)/μS/cm)、total dissolved solids/(mg/L)、phosphate ions/(mg/L)

3、Cross-flow and once-through steam boiler water quality:

boiler feed water:

turbidity/FTU、total hardness/(mmol/L)、pH(25℃)、dissolved oxygen/(mg/L)、oil/(mg/L)、iron/(mg/L) 、total alkalinity/(mmol/L)、phenolphthalein alkalinity/(mmol/L)、conductivity(25℃)/μS/cm) 、total dissolved solids/(mg/L)、phosphate ions/(mg/L)、sulfite/(mg/L)

boiler water:

total alkalinity/(mmol/L)、phenolphthalein alkalinity/(mmol/L)、pH(25℃)、conductivity(25℃)/μS/cm) 、total dissolved solids/(mg/L)、phosphate ions/(mg/L)、sulfite/(mg/L)

4、Back water quality:

total hardness/(mmol/L)、iron/(mg/L)、copper/(mg/L)、oil/(mg/L)

5、Hot water boiler water quality:

make-up water:

total hardness/(mmol/L)、pH(25℃)、turbidity/FTU、iron/(mg/L)、dissolved oxygen/(mg/L)、chlorine ions/(mg/L)

boiler water:

pH(25℃)、phosphate ions/(mg/L)、iron/(mg/L)、oil/(mg/L)、phenolphthalein alkalinity /(mmol/L)、dissolved oxygen/(mg/L)




Application Products
Classic Case
+8602985515255
yrhb@erunwas.com