Brand: Erun
Portable multi-parameter water quality tester
number: ERUN-SP7
Parameters: COD, ammonia nitrogen, total phosphorus, total nitrogen, turbidity, copper, etc.
Erun environmental protection ERUN-SP7 portable multi-parameter water quality detector, can according to the needs of heavy metals and inorganic salts (including COD, ammonia nitrogen, total phosphorus, total nitrogen, total phosphorus, total nitrogen, residual chlorine, total chlorine, chlorine dioxide, turbidity, color, copper, chromium, manganese, iron, zinc, aluminum, cyanide, phosphate, sulfate, nitrite, hardness, etc.), and the instrument can be expanded and customized test items according to the demand.
1. Repeatability: 3%
2. Optical stability: the light absorption value of the instrument drift is less than 0.002A in 20min
3. Reading mode: concentration, absorbance, and light transmittance
4. overall dimensions: 410mm 300mm 130mm
5. Weight: 4kg
6. Host power consumption: <10W
7.[计] measuring range
numbe | parameter | measurement rang | unit | error |
1 | COD (chromium method low range) | 5-200 | mg/L | 5% |
COD (Medium range) | 200-1000 | mg/L | 5% | |
COD (high range) | 1000-2000 | mg/L | 5% | |
2 | Ammonia nitrogen (low range) | 0.02-5.00 | mg/L | 3%(F.S) |
Ammonia nitrogen (high range) | 5.00-25.00 | mg/L | 3%(F.S) | |
3 | Total phosphorus(low range) | 0.02-2.00 | mg/L | 5%(F.S) |
Total phosphorus(high range) | 2.00-10.00 | mg/L | 5%(F.S) | |
4 | Total nitrogen (low range) | 0.10-15 | mg/L | 5%(F.S) |
Total nitrogen (high range) | 15-150 | mg/L | 5%(F.S) | |
5 | CODMn (Permanganate index) | 0.00-10.00 | mg/L | 0.2或5% |
6 | Residual chlorine (low range) | 0.02-2.00 | mg/L | 0.1或5% |
Residual chlorine (high range) | 2.00-10.00 | mg/L | 5% | |
7 | Total chlorine (low range) | 0.02-2.00 | mg/L | 0.1或5% |
Total chlorine (high range) | 2.00-10.00 | mg/L | 5% | |
8 | Chlorine dioxide (low range) | 0.05-2.00 | mg/L | 0.2或5% |
Chlorine dioxide (high range) | 2.00-20.00 | mg/L | 5% | |
9 | Sulfate | 5-200 | mg/L | 5% |
10 | Nitrate (nitrogen) | 0.02-10.0 | mg/L | 0.1或5% |
11 | Phosphate(low range) | 0.01-10.0 | mg/L | 0.2或5% |
Phosphate(high range) | 10.0-50.0 | mg/L | 5% | |
12 | Nitrite nitrogen(low range) | 0.02-1.0 | mg/L | 0.05或5% |
Nitrite nitrogen(high range) | 1.0-5.0 | mg/L | 5% | |
13 | Sulfides | 0.005-1.000 | mg/L | 5%(F.S) |
14 | Fluoride | 0.05-2.0 | mg/L | 5%(F.S) |
15 | Cyanide | 0.01-0.50 | mg/L | 5%(F.S) |
16 | Bromide | 0.10-5.00 | mg/L | 5%(F.S) |
17 | Formaldehyde | 0.05-2.5 | mg/L | 5%(F.S) |
18 | Volatile phenol (low range) | 0.20-5.0 | mg/L | 5%(F.S) |
Volatile phenol (high range) | 5.0-10.0 | mg/L | 5%(F.S) | |
19 | Hydrazine | 0.01-0.50 | mg/L | 5%(F.S) |
20 | Aniline compoun (low range) | 0.03-1.0 | mg/L | 5%(F.S) |
Aniline compoun(high range) | 1.0-5.0 | mg/L | 5%(F.S) | |
21 | Silica | 0.01-5.00 | mg/L | 5%(F.S) |
22 | Total alkalinity (low range) | 0-100 | mg/L | 5%(F.S) |
Total alkalinity (high range) | 100-1000 | mg/L | 5%(F.S) | |
23 | Total hardness (low range) | 0.1-5.0 | mg/L | 5%(F.S) |
Total hardness (high range) | 5.0-200.0 | mg/L | 5%(F.S) | |
24 | Chloride ion (chloride) | 0.20-100.0 | mg/L | 5%(F.S) |
25 | Copper (low range) | 0.02-1.0 | mg/L | 5%(F.S) |
Copper (high range) | 1.0-10.0 | mg/L | 5%(F.S) | |
26 | Zinc | 0.2-5.0 | mg/L | 5%(F.S) |
27 | Nickel (low range) | 0.02-0.50 | mg/L | 5%(F.S) |
Nickel (high range) | 0.10-10.0 | mg/L | 5%(F.S) | |
28 | Iron (low range) | 0.0-5.0 | mg/L | 5%(F.S) |
Iron (high range) | 5.0-10.0 | mg/L | 5%(F.S) | |
29 | Hexavalent chromium | 0.01-1.00 | mg/L | 5%(F.S) |
30 | Cadmium | 0.05-0.50 | mg/L | 5%(F.S) |
31 | Manganese (low range) | 0.01-0.10 | mg/L | 5%(F.S) |
Manganese (middle range) | 0.10-1.0 | mg/L | 5%(F.S) | |
Manganese (high range) | 0.10-5.00 | mg/L | 5%(F.S) | |
32 | Total chromium | 0.02-1.0 | mg/L | 5%(F.S) |
33 | Aluminum (low range) | 0.01-0.40 | mg/L | 5%(F.S) |
Aluminum (high range) | 0.40-4.00 | mg/L | 5%(F.S) | |
34 | Silver | 0.01-1.00 | mg/L | 5%(F.S) |
35 | Molybdenum | 0.05-1.00 | mg/L | 5%(F.S) |
36 | Arsenic (low range) | 0.004-0.100 | mg/L | 5%(F.S) |
Arsenic (high range) | 0.100-0.500 | mg/L | 5%(F.S) | |
37 | Beryllium | 0.001-0.10 | mg/L | 5%(F.S) |
38 | Boron | 0.10-2.00 | mg/L | 5%(F.S) |
39 | Mercury | 0.002-0.50 | mg/L | 5%(F.S) |
40 | Lead | 0.01-0.50 | mg/L | 5%(F.S) |
41 | Chroaticity (low range) | 5-50 | PCU | 3%(F.S) |
Chroaticity (high range) | 50-500 | PCU | 3%(F.S) | |
42 | Turbidity | 0-400 | NTU | ≤2%(F.S) |
43 | Suspended solids | 5-1000 | mg/L | ≤2%(F.S) |
44 | PH(Electrode method) | 0.00-14.00 | PH | ±0.01 |
45 | Dissolved oxygen (colorimetric method) | 0.1-10.0 | mg/L | 0.1±5% |
46 | Ozone | 0.01-2.5 | mg/L | 5%(F.S) |
47 | Anionic surfactants | 0.01-1.0 | mg/L | 5%(F.S) |
48 | Available chlorine | 0.0-150.0 | mg/L | 5%(F.S) |
49 | Chlorite | 0.00-20.00 | mg/L | 5%(F.S) |
50 | Stibonium (Sb) | 0.05-10.0 | mg/L | 5%(F.S) |
51 | Conductivity | 0.00µs/cm-199.9ms/cm | ms/cm | 1.0% (F.S) |
52 | Urea | 0.1-4.00 | mg/L | 8%(F.S) |
53 | Cyanuric acid | 5-200 | mg/L | 5%(F.S) |
54 | Salinity | 0-1000 | mg/L | 8% |
55 | Temperature | (0-99℃) | ℃ | ±1 |
56 | Ammonium ion | 0-25.00mg/L | mg/L | 3%(F.S) |
57 | Ferrous | 0.0-10.0 | mg/L | 5%(F.S) |
1. The wavelength is automatically selected when measuring the sample, without adjustment.
2. Adopt 5.0-inch large LCD screen, display in Chinese (available in English). The measured concentration is directly displayed and the absorbance is measured.
3. The instrument is convenient to carry around for on-site testing. The instrument has many functions, such as measurement, setting, recording, saving and data statistical processing.
4. With data storage function, it can store 100 curves and 1999 data with time and parameter labels. The built-in printer enables the direct printing of the measurement results.
5. Data interface: peripheral USB interface to support data management and instrument control.
6. Cold light source, narrow-band interference optical system, good optical stability. Optical path system: using solid light emitting device as both light source and as monochromatic, light source / monochromatic, color slot, sensor integrated, no moving parts, pulse power supply, the service life of the light source up to 100,000 hours.
7. Data power-off protection function.
8. Power supply mode: dual power supply system, built-in lithium battery, and has the power self-diagnosis function, automatically prompt the power is insufficient.
9. Online software: it can realize the input of the sample name, the tested items, testing institutions, inspection personnel and other information, and can export the text format data. Test report can be written directly, and report printing can be realized through the external printing system.
10. With special reagents, the determination of the measured object can be realized only with the zero point correction. The operation accuracy is high, simple and fast.
11. Expanded and customized test projects.