Water quality analyzer for fishery and aquaculture

The synopsis of project
The ERUN series water quality analyzer for fishery and aquaculture developed and produced by Erun Group is a solution designed with reference to the "GB 11607-1989". The water quality analyzer for aquaculture can mainly be used to test suspended solids, pH, dissolved oxygen, biochemical oxygen demand, total coliforms, mercury, cadmium, lead, hexavalent chrome, copper, zinc, nickel, arsenic, cyanide, sulfide, fluoride, ammonia nitrogen, volatile phenols, oil, etc. The water quality testing instrument for fisheries area is suitable for testing the water quality of marine and freshwater fishery waters such as spawning grounds, feeding grounds, wintering grounds, migration channels, and aquaculture areas for fish and shrimp.
The contents of project

1、Water quality standards for fishery and aquaculture:

Serial No.ItemsStandard values
1color, odor, and tasteDo not allow fish, shrimp, shellfish, or algae to have any color, odor, or peculiar smell
2Floating materialThere shall be no obvious oil film or floating foam on the water surface
3suspended solids
The artificially added amount shall not exceed 10, and suspended solids matter shall not have harmful effects on fish, shrimp, and shellfish after sedimentation at the bottom
4pHFreshwater 6.5~8.5, seawater 7.0~8.5
5

dissolved oxygen

In a continuous 24-hour period, the duration of 16 hours or more must be greater than 5 hours, and at any other time it must not be less than 3 hours. For salmon fish inhabiting frozen waters, the duration of any other time must not be less than 4 hours
6biochemical oxygen demand(five-day、20℃)Not exceeding 5, with a maximum ice period of 3
7total coliformsNot exceeding 5000/L (shellfish aquaculture water quality not exceeding 500/L)
8mercury≤ 0.0005
9cadmium≤ 0.005
10lead≤ 0.05
11hexavalent chrome≤ 0.1
12copper≤ 0.01
13zinc≤ 0.1
14nickel≤ 0.05
15arsenic≤ 0.05
16cyanide≤ 0.005
17sulfide≤ 0.2
18fluoride (calculated as F -)≤ 1
19ammonia nitrogen≤ 0.02
20kjeldahl nitrogen≤ 0.05
21volatile phenols≤ 0.005
22yellow phosphorus≤ 0.001
23oil≤ 0.05
24acrylonitrile≤ 0.5
25acrolein≤ 0.02
26hexachlorocyclohexane (C)≤ 0.002
27DDT≤ 0.001
28malathion≤ 0.005
29sodium pentachlorophenol≤ 0.01
30dimethoate≤ 0.1
31methamidophos≤ 1
32methyl-parathion≤ 0.0005
33
carbofuran≤ 0.01

Note: The unit is mg/L.


2、Technical parameters of water quality analyzer for fishery and aquaculture:

Serial No.

Items

Measuring range

Unit

Maximum deviation

Analysis methods

Standard basis

1

suspended solids

5-1000

mg/L

≤2%(F.S.)

Light scattering method

——

2

pH

0.00-14.00

——

±0.1

Electrode method

GB/T 6920

3

dissolved oxygen

0-20.0

mg/L

±0.2

Electrochemical probe method

GB/T 11913

4

biochemical oxygen demand

0-4000

mg/L

±8%

Dilution and seeding method

HJ 505

5

total coliforms

——

pcs/L

——

Multi tube fermentation method and membrane filtration method

GB 5750

6

mercury

0.002-0.50

mg/L

5%(F.S.)

Cold atomic absorption spectrophotometry

GB/T 7468

7

cadmium

0.05-0.50

mg/L

5%(F.S.)

Spectrophotometric method with dithizonc

GB 7471

8

lead

0.01-0.50

mg/L

5%(F.S.)

Atomic absorption spectrometry

GB 7475

9

hexavalent chrome

0.01-1.00

mg/L

5%(F.S.)

1,5 Dtphenylcarbohydrazide spectrophotometric method

GB 7467

10

copper

0.02-10.0

mg/L

5%(F.S.)

Sodium diethyldithiocarbamate spectrophotometric method

GB 7474

11

zinc

0.2-5.0

mg/L

5%(F.S.)

Dithizone spectrophotometric method

GB 7472

12

nickel

0.02-10.0

mg/L

5%(F.S.)

Dichloroxime spectrophotometric method

GB 11910

13

arsenic

0.004-0.500

mg/L

5%(F.S.)

Atomic fluorescence spectrometry

HJ 694

14

cyanide

0.01-0.50

mg/L

5%(F.S.)

Isonicotinic acid pyridine ketone colorimetric method

GB 7486

15

sulfide

0.005-1.000

mg/L

5%(F.S.)

Methylene blue spectrophotometric method

GB/T 16489

16

fluoride

0.05-2.0

mg/L

5%(F.S.)

Visual colorimetry with zirconium alizarinsulfonate

GB 7482

17

ammonia nitrogen

0.02-25.00

mg/L

3%(F.S.)

Nessler’s reagent spectrophotometry

GB 7479

18

volatile phenols

0.20-10.0

mg/L

5%(F.S.)

4-AAP spectrophotometric method

GB/T 7490

19

oil

0.00~900.00

mg/L

RSD<0.6%

Infrared photometric method

GB/T 16488


3、List of water quality analyzer for fishery and aquaculture:

Serial No.

Part Number

Name

Items

Pictures

1

ERUN-SP-M9000

Portable multi-parameter water quality tester

suspended solids, pH, dissolved oxygen, mercury, cadmium, lead, hexavalent chrome, copper, zinc, nickel, arsenic, cyanide, sulfide, fluoride, ammonia nitrogen, volatile phenols

Portable multi-parameter water quality tester for fishery and aquaculture


2

ERUN-BY-2A

Intelligent water quality BOD5 testing instrument

 biochemical oxygen demand

Intelligent water quality BOD5 testing instrument for fishery and aquaculture

3

ERUN-ST-OD380

Laboratory benchtop infrared spectrophotometer

oil

Laboratory benchtop infrared spectrophotometer for fishery and aquaculture

4

ERUN-SP-MD690

Portable water quality microbiological tester

total coliforms

Portable water quality microbiological tester for fishery and aquaculture

Note: Analysis parameter items can be flexibly combined and selected according to the customer's own needs.


4、Other parts list of water quality analyzer for fishery and aquaculture:

Serial No.

Name

Specifications

Quantity

1

colorimetric tube

Ф25×95mm

10

2

test tube rack

white

2

3

pipette gun

minimum graduation 0.1mL

1

4

cuvette

70×25×70mm

2

5

lens hood

black

1

6

mercury reagent50 times/set

1

7

cadmium reagent50 times/set

1

8

lead reagent50 times/set

1

9

hexavalent chrome reagent100 times/set

1

10

copper reagent

100 times/set

1

11

zinc reagent

50 times/set

1

12

nickel reagent

50 times/set

1

13

arsenic reagent50 times/set

1

14

cyanide reagent

50 times/set

1

15

sulfide reagent50 times/set

1

16

fluoride reagent

50 times/set

1

17

ammonia nitrogen reagent200 times/set

1

18

volatile phenols reagent100 times/set

1

19biochemical incubatorERUN-SHX1001
20temperature electrode/1
21dissolved oxygen electrode/1
22pH composite electrode/1
23chargers and charging cables/1
24USB cable/1
25instruction manual/1
26

product certificate

/product warranty card

/1

Note: Analysis parameter items can be flexibly combined and selected according to the customer's own needs.

Application Products
Classic Case
+86 18166600151
yrhb@erunwas.com